求翻译:This section describes how to calibrate the TEC 3000 temperature sensors. There are two calibration procedures: single point and two point calibration. For single point calibration, the reference point is LN2. For two point calibration, the reference points are LN2 and ice water. Unless regulations require a two p是什么意思?

待解决 悬赏分:1 - 离问题结束还有
This section describes how to calibrate the TEC 3000 temperature sensors. There are two calibration procedures: single point and two point calibration. For single point calibration, the reference point is LN2. For two point calibration, the reference points are LN2 and ice water. Unless regulations require a two p
问题补充:

  • 匿名
2013-05-23 12:21:38
本节介绍如何校准TEC 3000温度传感器。校准程序有两种:单点和两点校准。单点校准,参考点是液氮。二点校准,参考点是液氮和冰水。除非法规要求两点校正,单点校准程序的建议。两点校准的好处是更准确地测量温度接近室温环境。缺点是更长,更复杂的校准过程。单点校准的好处是一个简单的校准程序。单点校准的缺点是不准确的测量温度接近室温环境。